GENERAL ORDERING INFORMATION:
Specify Meter Color & Digit Size, Meter Power and a Signal Conditioner from the list below. Specify Analog Output, Setpoint Output and Digital Interface if needed. These meters are shipped with a standard configuration and range shown below for each signal conditioner and output option type. All information for making changes is covered in the supplied Owners Manual. For other scalings or configurations, add the SC (Special Configuration) option.
You must be logged in to post a review.
Micro Series meters with the temperature signal conditioner board provide highly accurate, stable and repeatable temperature indication for multiple thermocouple types and 100 Ohm platinum RTDs. The meters accept type J, K, T, E, N, R and S thermocouples plus 100 ohm platinum RTD’s with DIN alpha of 0.00385 or ANSI alpha of 0.003925. The thermocouple or RTD type, range and resolution (1° or 0.1°) are selectable from the front panel or via the serial interface. Display in Kelvin or Rankine can be selected by offsetting the Celsius or Fahrenheit range.
Cold junction compensation automatically corrects for temperature variations at the thermocouple reference junction at the meter. Open sensor indication is standard and may be set up to indicate either upscale or downscale. RTD excitation is provided by the meter. RTD connections can be of the 2-, 3- or 4-wire type. For 3- and 4-wire connections, the meter automatically compensates for lead resistance of the sensor.
All ranges for all temperature sensor types are digitally calibrated at the factory. This eliminates the need for field calibration or drift caused by potentiometers found in non-microcomputer based meters.
All Micro Ssries meters allow exceptionally fast read rates. Concurrent Slope (Pat 5,262,780) is a method of analog-to-digital conversion that allows up to 60 conversions per second while integrating the input over a full power cycle. Fast read rates provide true readings and fast response
Plug-in isolated analog output, dual-setpoint controller, RS232/485 serial communicationsboards can upgrade Micro and Mighty Series meters from stand-alone monitor to system interface and control. A Micro Series temperature meter can serve as a linearized, scalable, ighly accurate, isolated, 4-20 mA transmitter.
Micro Series meters can be powered from 95-240 Vac ±10% and 95-370 Vdc, or optionally from 12-34 Vac and 10-48 Vdc. Any setup functions and front panel keys may be locked out for simplified usage and security. The popular 1/8 DIN version with 0.56″ high digits is environmentally sealed to NEMA-4X (IP65) from the front when panel mounted, and is available with red or green LEDs.
| TC Type | Range | Conformity Error |
| J | -210°C to +760°C -347°F to +1400°F | 0.09°C 0.16°F |
| K | -244°C to +1372°C -408°F to +2501°F | 0.1°C 0.17°F |
| T | 0°C to +400°C -257°C to 0°C +32°F to 752°F -430°F to +32°F | 0.03°C 0.2°C 0.05°F 0.36°F |
| E | -240°C to +1000°C -400°F to +1830°F | 0.18°C 0.32°F |
| N | -245°C to +1300°C -410°F to +2370°F | 0.10°C 0.17°F |
| R | -45°C to +1768°C -49°F to +3214°F | 0.17°C 0.31°F |
| S | -46°C to +1768°C -51°F to +3213°F | 0.12°C 0.22°F |
| Applicable to all TC types | |
| Calibration | NIST Monograph 125 (IPTS-68) |
| Input resistance | 1 Gohm |
| Input current | 100 pA |
| Max lead resistance | 1 Gohm max for rated accuracy |
| Span tempco | ±0.003% of reading/ºC |
| Reference junction tempco | ±0.02 deg/deg |
| Span tempco | ±0.003% of reading/ºC |
| Overvoltage protection | 125 Vac |
| NMR at 50/60 Hz | 80 dB plus selectable digital filter from 80 ms |
| to 9.6 s time constant | |
| CMR, DC-60 Hz | 120 dB with 500 W imbalance |
| CMV, DC-60 Hz | 250 Vac from power and earth grounds |
| Open sensor indication | Flashes full-scale |
| RTD Alpha | Range | Conformity Error |
| DIN .00385 | -202°C to +850°C -331°F to +1562°F | 0.03°C 0.05°F |
| ANSI .003925 | -202°C to +631°C -331°F to +1168°F | 0.04°C 0.07°F |
| Applicable to all RTD types | |
| Calibration, DIN | IEC 751 (IPTS-68) |
| Calibration ANSI | NIST Monograph 126 |
| Configuration | 2, 3 or 4-wire connection |
| Excitation current | 0.2 mA |
| Span tempco | ±0.003% of reading/ºC |
| Zero tempco | ±0.03 deg/deg |
| Sensor lead resistance | 2-wire, 10 mdeg/W/deg up to 10 W |
| tempco per conductor | 3 & 4-wire, 10 mdeg/W/deg up to 100 W |
| Overvoltage protection | 125 Vac |
| Open sensor indication | Flashes full-scale |
| Display | |
| Readout | 5 digits, 7-segment |
| Digit Height | Available with 0.56", 1.0", 4.0" or 5.0" digits. |
| Color | High brightness red. Green available with 0.56" digits. |
| Indicators | Minus sign, 2 red LED lamps |
| A-to-D Conversion | |
| Technique | Concurrent Slope™ (Pat 5,262,780) |
| A-to-D Rate | 60/s at 60 Hz, 50/s at 50 Hz |
| Output Update | 56/s at 60 Hz, 47/s at 50 Hz |
| Display Update | 3.5/s at 60 Hz, 3/s at 50 Hz |
| Power | |
| Voltage, std. | 95-240 Vac ±10% and 95-370 Vdc |
| Voltage, opt. | 12-34 Vac and 10-48 Vdc |
| Frequency | DC or 49-440 Hz |
| Power isolation | Safety-rated to 250 Vac, meter ground to earth ground, |
| DC to 60 Hz, 4.2 kVp per High Voltage Test |
| Environmental | |
| Operating Temperature | 0°C to 60°C |
| Storage Temp. | -40°C to 85°C |
| Relative Humidity | 95% at 40°C, non-condensing |
| Protection | NEMA-4X (IP-65) when panel mounted |
| Operating Temperature |
The Micro temperature meter allows 2-, 3- and 4-wire RTD hookup to the J5 connector. The meter applies an excitation current of 0.2 mA, which it monitors to make ratiometric corrections for excitation supply variations.
With the optional analog output board, Micro Series temperature meters can serve as superb, isolated 4-20 mA transmitters. The analog output is scaled to the display, which is linearized to ºC or ºF and is exceptionally accurate. The analog output further tracks the high read rate of the meter, at up to 60 readings per second at 60 Hz power. Fast update rates are beneficial in many closed-loop and PID control applications.
With the optional dual contact relay or dual solid state relay output board options, Micro Series temperature meters can serve as extremely fast and accurate ON/OFF controllers for closed-loop temperature control.
Multiple setpoint operating modes are individually selectable for each relay, as explained in the dual-setpoint controller section. Relay duty cycles and chatter can be minimized with programmable hysteresis and time delays. High duty cycles and extremely fast response times are possible with the solid state relay, which has a typical response time of only 17 ms.
By using the optional dual contact relay or dual solid state relay output options, Micro Series temperature meters can monitor processes and provide alarms or shutoffs when these processes exceed normal limits.
A band deviation operating mode can be selected for each relay, where an alarm is generated whenever the reading is a selected number of counts above or below the setpoint. Relay operation can be selected as latching or non-latching. When an alarm or shutdown condition is reached, a latched output will remain in the alarm condition until it is reset by a front panel pushbutton, via the serial interface, or via the rear connector.
© 2024 Electro-Numerics. All Rights Reserved. Privacy Policy. Website by Mathis Consulting LLC | Developed by Cebu Web Maker